Specifications

Phase 3 – Performance Specifications

 

Tests Performed

Defibrillator Energy, Charge Time and Cardioversion Automated External Defibrillator (AED) Performance
ECG Monitor Performance
Pacer Sensitivity, Refractory Periods and Noise Immunity
Pacer Pulse Characteristics
Energy Measurement, General:
Load resistance: 50 ohms ±1%, non-inductive
ECG amplitude at defib pads: 1 mV QRS
WAVEFORM (oscilloscope) Output
High Range: 1000:1 amplitude attenuation
Low Range: 200:1 amplitude attenuation
Waveform Playback: 200: 1 time base expansion
First Phase Pulse Width: 0.1 – 58 msec
Second Phase Pulse Width: 0.1 – 58 msec
Interphase Delay: 0.1 – 58 msec
Tilt: 0 – 99.9%
Modulation Frequency: 1000 – 8000 Hz
Modulation Duty Cycle: 0 – 99.9%
Test Pulse: 46 Joules ±10%
Defibrillator High Range Energy Test:
Energy Measurement: 0.0 to 600.0 Joules (±1% ±2 LSD)
Voltage Measurement: 0 to 5000 Volts (±1% ±2 LSD)
Current Measurement: 0.0 to 100.0 Amps (±1% ±2 LSD)
Pulse Width Measurement: 0.5 to 58.36 msec (±1% ±2 LSD)
Trigger Level: 80 Volts
Playback Amplitude: 1 mV per 1000 volts on Lead II;
1 mV per 2000 volts at defib pads
Test Pulse: 126 Joules ±10%
Defibrillator Low Range Energy Test:
Energy Measurement: 0.0 to 50.0 Joules (±1% ±2 LSD)
Voltage Measurement: 0 to 1000 Volts (±1% ±2 LSD)
Current Measurement: 0.0 to 20.0 Amps (±1% ±2 LSD)
Pulse Width Measurement: 0.5 to 58.36 msec. (±1% ±2 LSD)
Trigger Level: 16 Volts
Playback Amplitude: 1 mV per 200 volts on Lead II;
1 mV per 400 volts at defib paddles
Defibrillator Charge Time Test:
Charge Time Measurement: 0.0 to 99.9 seconds (±1 LSD)
Defibrillator Cardioversion Test:
Sync Delay Measurement: -200 to +800 msec. (±1 LSD)
Delay Target: 20 to 65 msec window when enabled
Sync Point: Selectable, peak of ECG Q or R wave
AED Performance Test:
Test Method: Verify AED shock advisory for specified arrhythmia
Pacemaker Pulse Test:
Pulse Amplitude Measurement: 4 to 250 milliamps, all loads (±1% ±1 LSD)
Pulse Rate Measurement: 20 to 220 PPM (±1% ±1 LSD)
Pulse Width Measurement: 0.5 to 58.36 msec. (±1% ±2 LSD)
Test Load Range: 50 to 1600 ohms, in 50 ohm steps
WAVEFORM output: 50 milliamps per volt, all loads
Measurement Methods: Average, leading edge, trailing edge, peak
Test Pulse: 145 mA ±10%
Pacemaker Noise Immunity Test:
Test Waveform: 50Hz or 60Hz sine wave
Noise Amplitude Range: 0.18 to 10.00 mV peak-to-peak
Noise Amplitude Resolution: 0.139 mV
Pacemaker Sensitivity Test:
Test Waveform: Square (SQR), Triangle (TRI) or Haversine (SSQ) pulse
Waveform Width: 10, 25, 40, 100, or 200 msec.
Amplitude Range: 0.00 to 3.00 mV peak
Pacemaker Refractory Period Test:
Paced Refractory Period (PRP): 50 to 750 msec. (±1 LSD)
Sensed Refractory Period (SRP): 50 to 750 msec. (±1 LSD)

ECG Simulator

Performance Test Waveforms:
DC Pulse, 4 seconds, Square Wave, 2 Hz, Triangle Wave, 2 Hz Sine Wave @ 0.1, 0.5, 10, 20, 40, 50, 60, 70, or 100 Hz
Normal Sinus Rhythm:
30, 60, 90, 120, 150, 180, 240 or 300 BPM
Cardioversion, Shock Advisory and AED Test Waveforms:
Atrial Fibrillation, Coarse
Atrial Fibrillation, Fine
Asystole 1 (random, low-frequency baseline fluctuation)
Asystole 2 (flat line/zero volts) Supraventricular Tachycardia (SVT-140)
VTACH @ 140, 160, 190 BPM Torsades de Pointe @ 200 BPM
Coarse Ventricular Fibrillation (CVF) / Fine Ventricular Fibrillation (FVF)
Square Pulse 1ms to 60 BPM
Arrhythmia Simulations:
Second Degree A-V Block, Premature Atrial Contraction (PAC), R-on-T PVC
Right Bundle Branch Block (RBBB), Premature Ventricular Contraction (PVC)
Multifocal PVC, Run of 5 PVC, Bigeminy, Trigeminy
Pacemaker Test Waveforms:
SQR (square) Pacer Trigger, width = 2, 25, 40, 100 or 200 msec
TRI (triangle) Pacer Trigger, width = 10, 25, 40, 100 or 200 msec
SSQ (haversine) Pacer Trigger, width = 10, 25, 40, 100 or 200 msec
Performance Specifications:
Output Level: Selectable, 1 mV, 2 mV or 0.5 mV into ECG Lead II
Impedance: 500 ohms (±0.2%), Amplitude: ±2%

 

Non-Volatile Memory

Data Capacity:
50 Test Records, 10 Defibrillator Waveform Records, 32 Autosequences
Test Record Content:
Device ID, Time/date of test, Test type (Manual or Auto), Device type (defib. or AED)
Up to 10 defibrillator energy tests (or 32 AED energy tests)
1 defibrillator charge time test
Up to 4 defibrillator cardioversion tests
Up to 12 ECG performance tests
Up to 10 pacer pulse tests
1 pacer noise immunity test
Up to 2 pacer sensitivity tests
Up to 2 pacer refractory period tests
Interface:
LCD (5.2″ x 1.5″; 40 characters x 8 lines text; 240 x 64 pixel graphics) Defibrillator Input: Molex 42820-3212
Pacemaker Input: 2 x safety-style banana jack (red (+)/black (-))
ECG Simulator Outputs: 10 x safety banana jack (RA; RL; LA; LL; V1-V6)
Defibrillator/Pacer Waveform Output: 1/8″ mono phono jack
High-Level ECG Output: 1/8″ mono phono jack
USB Port: Type “B”, USB 1.1 or USB 2.0 compatible.
Serial (RS-232) Port: DB9 Male, RS-232C, bi-dir, CTS handshaking, 9600 baud, 8-N-1 Keyboard Port: PS/2 (6-pin miniDIN female)
Power Supply:
Internal 12.5V/1.4A-h NiCad, 24 hours of use between charges
Environment
15°C to 40°C, 10% to 90% RH, Indoor Use Only, Category II
Dimensions
9.5″ W x 8″ H x 5.5″ D (24cm W x 20cm H x 14cm D)
Weight
3 lbs. (1.4 kg)

 