Specifications

Incutest – Performance Specifications

Temperature:
Measurement Range: 0° C to 50° C
Accuracy: T1+/- 0.05° C
Resolution: 0.01° C
Convection: 5 “puck” sensors (T1-T5) – radiant warmer
Air Convection: 5 sensors (alternate “pos”’ T1 – T5) – incubator
1 Mattress Temperature Sensor 
Skin Sensor²: Sensor “oven” with thermal pad contact 
28° C to 40° C in 2° C increments 
Resolution: 0.01° C 
Accuracy: +/- 0.05° C 
Airflow:
Measurement Range: 0.1 to 1 m/s
Accuracy: +/- 0.1 m/s at 50% RH +/- 15%
Resolution: 0.01m/s
Sound Level:
Measurement Range: 30 dbA to 90 dbA, Type II
Accuracy: +/- 5 dbA
Resolution: 0.1 dbA
Relative Humidity:
Measurement Range: 0 – 100% Non-condensing
Accuracy: +/- 3 % to 100% RH
Resolution: 0.1% RH
Data Retention:
Capacity: up to 48 hours per test
Sample rate: 1/min or 1/sec, automatic, parameter dependent 

Test Specifications:

Incubator Testing Specifications
  • • Measures temperature at 6 points as per IEC 60601‐2‐19 (sub‐clause 50.102)
  • • Measures humidity at the center of the enclosure per IEC 60601‐2‐19 (sub‐clause 103.1)
  • • Measures airflow at 4 points in the enclosure per IEC 60601‐2‐19 (sub‐clause 104.1)
  • • Measures sound level inside the incubator per IEC 60601‐2‐19 (sub‐clause 102.1)
  • • Measures sound level inside the incubator during alarm per IEC 60601‐2‐19 (sub‐clause 102.2)
  • • Measures sound level outside the incubator per IEC 60601‐2‐19 (sub‐clause 102.3)
  • • Measures incubator temperature at center point (A) of incubator as per IEC 60601‐2‐19 (subclause 50.102)
  • • Measures average incubator temperature (AIT) per IEC 60601‐2‐19 (sub‐clause 2.9.106)
  • • Measures incubator temperature (IT) per IEC 60601‐2‐19 (sub‐clause 2.9.105)
  • • Measures average temperature (AT) per IEC 60601‐2‐19 (subclause 2.9.103)
  • • Measures average temperature at 4 points per IEC 60601‐2‐19 (sub‐clause 50.102)
  • • Measures skin temperature sensor accuracy per IEC 60601‐2‐19 (sub‐clause 50.104)¹
  • • Confirms skin temperature sensor control of incubator per IEC 60601‐2‐19 (sub‐clause 50.105)
  • • Confirms independent incubator temperature sensor accuracy per IEC 60601‐2‐19 (sub‐clause 50.106)
  • • Measures difference between average temperature and control temperature per IEC 60601‐2‐19 (sub‐clause 50.107)
  • • Measures actual warm‐up time compared to operating manual per IEC 60601‐2‐19 (sub‐clause 50.108)
  • • Measures overshoot and recovery to steady state per IEC 60601‐2‐19 (sub‐clause 50.109)
  • • Confirms humidity reading per IEC 60601‐2‐19 (sub‐clause 50.110)
Radiant Warmer Testing Specifications
  • • Measures skin temperature sensor accuracy per IEC 60601‐2‐21 (sub‐clause 50.101)¹
  • • Measures average warmer temperature at 4 points per IEC 60601‐2‐21 (sub‐clause 50.102)
  • • Measures average warmer temperature at center point (A) of incubator as per IEC 60601‐2‐21(sub‐clause 50.102)
  • • Confirms skin temperature sensor control of warmer per IEC 60601‐2‐21 (sub‐clause 50.103)
  • • Confirms airflow is less than 0.1 m/s per IEC 60601‐2‐21 (subclause 50.102)
  • • Measures sound level on the mattress per IEC 60601‐2‐21 (sub‐clause 102.1)
  • • Measures sound level on the mattress during alarm per IEC 60601‐2‐21 (sub‐clause 102.1)
  • • Measures sound level in front of the warmer per IEC 60601‐2‐21 (sub‐clause 102.1)
Transport Incubators
Testing carried out in accordance with IEC601‐2‐20.
Utility App
Enables ad‐hoc testing of airflow, humidity, sound & temperature.

 

¹Sensor accuracy heater is an optional accessory.

All specifications subject to change without notice.