Specifications

(Infutest 2000 RSM – Performance Specifications)

Single Rate Test (Continuous Flow Conditions)

Flow Measurement
Nominal range: 0.1 – 1200 ml/hr
Max. flow measurable and displayed: 1700 ml/hr
Min. flow measurable and displayed: 0.04 ml/hr
Max. display resolution: 0.001 ml/hr
Measuring time: 10 min. at 0.1 ml/hr / 20 sec. max. above 6 ml/hr
Accuracy: Average flow: +/- 1%, 0.1- 1200 ml/hr
Ranges: LOW – 0.1 to 170 ml/h
HIGH – 170 to 1200 ml/h
Range selection: Automatic
Internal nominal effective collection volume: LOW range: 0.014 ml
HIGH range: 1.1 ml
Volume Measurement
Range: 0 – 9999 ml, Max. display resolution: 0.001 ml
Accuracy: +/- 1%
Back Pressure Measurement
Range: 0 – 300 mmHg, Zero offset error:+/- 5 mmHg,
Display resolution: 1 mmHg, Accuracy:+/- 1% FS, +/- zero offset error
Elapsed Time Measurement
Range: 0 – 100 hours, Display format:hh:mm:ss, Accuracy: +0/-1second

 

Pulsatile Flow Conditions

Flow Measurement

Nominal range: 5 – 999 ml/hr
Max. flow measurable and displayed: 1700 ml/hr
Min. flow measurable and displayed: 2.75 ml/hr
Max. display resolution: 0.001 ml/hr
Measuring time: 14 min. at 5 ml/hr / 20 sec. max. above 200 ml/hr
Accuracy: Average flow +/- 1%
Range: HIGH range only (internally selected)
Internal nominal effective collection volume: 1.1 ml
Volume Measurement
Range: 0 – 9999 ml, Max. display resolution: 0.001 ml,
Accuracy: +/- 1%

 

Dual Rate Test (Continuous Flow Conditions ONLY)

Flow Measurement
Nominal range: 0.1 – 170 ml/hr
Max. flow measurable and displayed: 200 ml/hr
Min. flow measurable and displayed: 0.04 ml/hr
Max. display resolution: 0.001 ml/hr
Measuring time: 10 min. at 0.1 ml/hr / 20 sec. max.
above 6 ml/hr
Accuracy: Average flow: +/- 1%
Ranges: LOW range only
Delivery period determination: Automatic
Internal nominal effective collection volume: 0.014 ml

 

PCA Pump Test (Continuous Flow Conditions ONLY)

Flow Measurement, Bolus Delivery period
Nominal range: 0.1 – 170 ml/hr
Max. flow measurable and displayed: 200 ml/hr
Min. flow measurable and displayed: 0.04 ml/hr
Max. display resolution: 0.001 ml/hr
Measuring time: 10 min. at 0.1 ml/hr / 20 sec. max. above 6 ml/hr
Accuracy: Average flow: +/- 1%
Ranges: LOW range only
Lockout detection: Automatic
Internal nominal effective collection volume: 0.014 ml
Occlusion Pressure Test
Display Units: mmHg, psi / Display resolution:1 mmHg(0.1 psi)
Range: 0 – 2586 mmHg(0 – 50 psi)
Measuring time: 2 seconds
Zero offset error: +/- 5 mmHg (0.1 psi) Accuracy:+/- 1% +/- zero offset error
Lockout Time Measurement
Range: 0 – 100 minutes, Display format:mm:ss,
Accuracy: +/- 2 seconds
Nurse Call Detection
Nurse Call signal from pump sampled once per second.
Nurse Call signal can be produced by any of: relay contacts, open collector output, TTL output, RS-232 level, or 20 mA current loop.
Memory Type:
NVRAM (battery-backed)
Capacity:
2 channel: 900 flow and 700 pressure measurements per channel
4 channel: 450 flow and 350 pressure measurements per channel
Interface
User Interface: 40-character by 8-line backlit LCD, 4 front
panel soft-keys,
Internal beeper: LCD contrast control
Fluid interface:
Inputs A & B: Delrin twistlock, self-sealing / Common output A & B: Delrin twistlock
PCA Pump Trigger Outputs:
Mechanical One 1/4″ female stereo phono jack per channel
Electrical Relay contacts rated at 120 VAC, 1 A; normally open
and normally closed contacts available.
Nurse Call Inputs:
Mechanical- One 1/4″ female stereo phono jack per channel
Electrical- 50K-ohm differential input impedance.
Parallel Port
Centronix standard printer interface, DB-25F
Printer Driver
Epson MX/FX Series, 80 cps ,80 columns and 66 lines / page
Serial Port:
Mechanical – RS-232 / DB-25M
Electrical – RS-232C; bidirectional; CTS handshaking;
9600 baud, 8-None-1
Power Supply:
120 VAC 60 Hz @ 12 W (230/240 VAC 50 Hz
Europe/UK)
Environment:
15°C to 40°C, 10% to 90% RH, Indoor Use Only, Category II;
Dimensions:
12″ W x 12″ D x 6″ H (30.5 cm W x 30.5 cm D x 15.2 cm H)
Weight:
10 lbs. (4.5 kg)

 

All specifications subject to change without notice.