Infutest 2000 Series E

Slide background
Slide background

Infutest 2000 Series E

Dual-Channel Infusion Device Analyzer

  • Test all types – dual-rate, pulsatile flow,and PCA infusion pumps
  • Analyze up to four channels simultaneously
  • Unique “Synchro-Start” function – optimize IV Pump and Analyzer accuracy

  • Overview
  • Features
  • Specifications
  • Accessories
  • The best solution for infusion device inspection
    Simultaneously inspect up to four single channel IV pumps or four independent channels of a multi-channel device in just seconds with our Infutest 2000 Series E and optional RSM. Better than 1% accuracy on all pumps, including Baxter Colleague and Alaris® Signature.

    Conduct the full range of flow, pressure, volume and interval tests on practically any fluid delivery system including single rate, dual rate and patient controlled analgesia (PCA) devices. These tests are performed following easy-to-use manual or autosequence formats using the front panel keypad and graphical display.

    Work smarter
    Interface Infutest 2000E to our ES601Plus Automated Safety Analyzer or other test systems such as a medTester 5000C with IVPUMP autosequence option.

    Add Infutest RSM for immediate upgrade to a four-channel analyzer. No need to send your Infutest in for a factory upgrade.
  • End-of-infusion analysis (EIA)
    Accurate volume-to-be-infused (VTBI) measurements are a snap. EIA calculates the volume infused at the instant the STOP key is pressed or when the autosequence preset test time has elapsed.

    SynchroStart
    Accurately captures the initial start-up flow dynamics of your IV pump, an invaluable feature when testing pumps with a pulsatile flow pattern. No need to attempt to start both the IV pump and the analyzer at the same time to make this critical measurement. Infutest 2000 does it for you...automatically!

    DTP-5 COMPANION SOFTWARE
    Infutest 2000 comes complete with Windows (98/2000/XP) based data transfer program.
    - (DTP-5) for quick and easy data transfer, graphical analysis, autosequence editing, plus organization and printing of test results.
  • Infutest 2000 Series E – Performance Specifications

    Single Rate Test (Continuous Flow Conditions)
    Flow Measurement
    Nominal range:0.1 - 1200 ml/hr
    Max. flow measurable and displayed:1700 ml/hr
    Min. flow measurable and displayed:0.04 ml/hr
    Max. display resolution:0.001 ml/hr
    Measuring time:10 min. at 0.1 ml/hr / 20 sec. max. above 6 ml/hr
    Accuracy: Average flow:+/- 1%, 0.1- 1200 ml/hr
    Ranges:LOW - 0.1 to 170 ml/h
    HIGH - 170 to 1200 ml/h
    Range selection:Automatic
    Internal nominal effective collection volume:LOW range: 0.014 ml
    HIGH range: 1.1 ml
    Volume Measurement
    Range:0 - 9999 ml, Max. display resolution: 0.001 ml
    Accuracy: +/- 1%
    Back Pressure Measurement
    Range:0 - 300 mmHg, Zero offset error:+/- 5 mmHg,
    Display resolution:1 mmHg, Accuracy:+/- 1% FS, +/- zero offset error
    Elapsed Time Measurement
    Range:0 - 100 hours, Display format:hh:mm:ss, Accuracy: +0/-1second
     
    Pulsatile Flow Conditions
    Flow Measurement
    Nominal range:5 - 999 ml/hr
    Max. flow measurable and displayed:1700 ml/hr
    Min. flow measurable and displayed:2.75 ml/hr
    Max. display resolution:0.001 ml/hr
    Measuring time:14 min. at 5 ml/hr / 20 sec. max. above 200 ml/hr
    Accuracy:Average flow +/- 1%
    Range:HIGH range only (internally selected)
    Internal nominal effective collection volume:1.1 ml
    Volume Measurement
    Range: 0 - 9999 ml, Max. display resolution: 0.001 ml,
    Accuracy: +/- 1%
     
    Dual Rate Test (Continuous Flow Conditions ONLY)
    Flow Measurement
    Nominal range:0.1 - 170 ml/hr
    Max. flow measurable and displayed:200 ml/hr
    Min. flow measurable and displayed:0.04 ml/hr
    Max. display resolution:0.001 ml/hr
    Measuring time:10 min. at 0.1 ml/hr / 20 sec. max.
    above 6 ml/hr
    Accuracy:Average flow: +/- 1%
    Ranges:LOW range only
    Delivery period determination:Automatic
    Internal nominal effective collection volume:0.014 ml
     
    PCA Pump Test (Continuous Flow Conditions ONLY)
    Flow Measurement, Bolus Delivery period
    Nominal range:0.1 - 170 ml/hr
    Max. flow measurable and displayed:200 ml/hr
    Min. flow measurable and displayed:0.04 ml/hr
    Max. display resolution:0.001 ml/hr
    Measuring time:10 min. at 0.1 ml/hr / 20 sec. max. above 6 ml/hr
    Accuracy:Average flow: +/- 1%
    Ranges:LOW range only
    Lockout detection:Automatic
    Internal nominal effective collection volume:0.014 ml
    Occlusion Pressure Test
    Display Units:mmHg, psi / Display resolution:1 mmHg(0.1 psi)
    Range:0 - 2586 mmHg(0 - 50 psi)
    Measuring time:2 seconds
    Zero offset error:+/- 5 mmHg (0.1 psi) Accuracy:+/- 1% +/- zero offset error
    Lockout Time Measurement
    Range:0 - 100 minutes, Display format:mm:ss,
    Accuracy:+/- 2 seconds
    Nurse Call Detection
    Nurse Call signal from pump sampled once per second.
    Nurse Call signal can be produced by any of: relay contacts, open collector output, TTL output, RS-232 level, or 20 mA current loop.
    Memory Type:
    NVRAM (battery-backed)
    Capacity:
    2 channel:900 flow and 700 pressure measurements per channel
    4 channel:450 flow and 350 pressure measurements per channel
    Interface
    User Interface:40-character by 8-line backlit LCD, 4 front
    panel soft-keys,
    Internal beeper:LCD contrast control
    Fluid interface:
    Inputs A & B:Delrin twistlock, self-sealing / Common output A & B: Delrin twistlock
    PCA Pump Trigger Outputs:
    MechanicalOne 1/4" female stereo phono jack per channel
    ElectricalRelay contacts rated at 120 VAC, 1 A; normally open
    and normally closed contacts available.
    Nurse Call Inputs:
    Mechanical- One 1/4" female stereo phono jack per channel
    Electrical- 50K-ohm differential input impedance.
    Parallel Port
    Centronix standard printer interface, DB-25F
    Printer Driver
    Epson MX/FX Series, 80 cps ,80 columns and 66 lines / page
    Serial Port:
    Mechanical - RS-232 / DB-25M
    Electrical - RS-232C; bidirectional; CTS handshaking;
    9600 baud, 8-None-1
    Power Supply:
    120 VAC 60 Hz @ 12 W (230/240 VAC 50 Hz
    Europe/UK)
    Environment:
    15°C to 40°C, 10% to 90% RH, Indoor Use Only, Category II;
    Dimensions:
    12" W x 12" D x 6" H (30.5 cm W x 30.5 cm D x 15.2 cm H)
    Weight:
    10 lbs. (4.5 kg)
     
    All specifications subject to change without notice.
  • Infutest 2000 Accessories

    3000-004Power Cord, UK
    3000-005Power Cord, Aus / CN
    3000-006Power Cord, EU
    3000-471Power Cord, NA
    3000-472Power Supply, SPS
    3140-040Cable, Serial Adapter, DB9F - DB25F (Infutest 2000 to PC via RS-232)
    3140-399Adaptor, USB To Serial DB9M
    7006-006Syringe, 60 cc, Luer Lock Tip
    7006-050Accessory Bag, 8.5" x 6" x 2.75" (Large)
    7006-051Accessory Bag, 8.5" x 2.9" x 2.6" (Small)
    7006-052Accessory Bag, 7" x 5" x 3" (Medium)
    7006-054Soft Case, Medium Multi-Product, 17" x 14" x 7"
    7006-055Soft Case, Large Multi-Product, 17" x 14" x 11"
    7100-060Cable Assy, NCA-1 (Infutest 2000 to Nurse Call Alarm)
    7100-061Cable Assy, PCA-1 (Infutest 2000 to Patient Controlled Analgesic)
    7100-062Cable Assy, SSC-1, DB25 - DB25 (Infutest 2000 to PC via RS-232)
    7100-064Cable Assy, PPC-1 Parallel Printer (Infutest 2000 to Centronix printer)
    7100-279Cable Assembly, AUX-1
    7300-005Assy, 2 CH IFT Input Output Tubing
    7500-087Kit, Factory/SC IFT Upgrade, D TO E - English
    7500-437Kit, Serial Adapter, ES601 -> Infutest 2000 (RJ12 -> DB25F) (ES601 to Infutest 2000)
    7500-439Kit, Serial Adapter, IFT 2000 TO PC Via USB (Infutest 2000 to PC via USB)
    7600-051Barcode CCD Scanner (RS232) c/w Case
    7600-052Barcode CCD Scanner (PS-2) Cc/wCase

    For more information, view the accessories list page

Infutest 2000 Series E
Product Sheet
Infutest 2000 Series E
Support
Find Out More