vPad-BP

NIBP Simulator

vPad-BP, part of the vPad-A1 family, is a NIBP (non invasive blood pressure) simulator designed to test the accuracy and repeatability of the NIBP function on Patient Monitors. User programmable settings allow for easy on site configuration to accommodate new models of NIBP devices.

  • Features
  • Specifications
  • Accessories
    • Compatible with all major industry manufacturers
    • Field editable Envelopes allows for easy set up of new patient monitors
    • Unlimited preset NIBP simulations
    • Easy to navigate
    • Auto Settings
    • Auto Sequences
    • Test reports
    • Field upgradeable
    • Control via Bluetooth or USB
  • vPad-BP™ Performance Specifications

    Manufacturer Envelopes
    • GE Dinamap and Dinamap Pro
    • Critikon Dinamap Plus
    • Datascope Passport
    • Welch-Allyn Spot LXi and 52000
    • Fukuda Dynascope
    • Colin Press-Mate
    • ...and more custom simulations
    Pressure Units
    mmHg, mbar, kPa, inH₂O, cmH₂O
    BP Presets
    Systolic/Diastolic (mmHg)
    Adult Neonatal
    255/195 150/100
    200/150 120/80
    150/100 100/65
    120/80 80/50
    100/65 60/30
    80/50 35/15
    60/30
    BP Simulation:
    Simulation type: oscillometric
    Rate Range: 20 – 240 BPM
    Accuracy:
    • ±0.25 BPM in sync mode
    • ±1 BPM otherwise
    Amplitude:
    • 0 – 2 mL
    • 1.25 mmHg into 500ml cuff
    Amplitude Range: 0 – 150%
    Amplitude Accuracy: better than 0.5%
    AutoSettings: unlimited, user definable
    Envelope Shift:
    ± 50 mmHg max
    Minimum Diastolic: 15 mmHg
    Maximum Systolic: 275 mmHg
    Manometer:
    Pressure Range: 0.0 to 400.0 mmHg
    Accuracy: ± 0.5 mmHg
    Resolution: 0.1 mmHg
    Regulated Pressure Source:
    Pressure Range: 10.0 to 400.0 mmHg
    Accuracy: ± 0.5 mmHg
    Resolution: 0.1 mmHg
    Leak Test:
    Automatic/manual Inflation
    Automatic Timer
    Leak Test Time: 30 - 600 seconds (user defined)
    Target Pressure: 20 to 400 mmHg
    Range: 0 to 200 mmHg/min
    User Definable Presets: 12
    User Definable AutoSettings: unlimited
    OverPressure Test:
    Automatic/manual Inflation
    Range: 20 - 400 mmHg
    Release Time: 1-999 sec
    User Definable Presets: 12
    User Definable AutoSettings: unlimited
    Standard Features / Accessories:
    • Autosequences
    • Unlimited User Defined Settings
    • vPad-A1 Power Base / Display
    • Universal Hose Adapter Kit
    Communication / User Interface
    via vPad-A1 Base Unit: Android 5" tablet:
    • Touchscreen User Interface
    • Wired (USB) or Bluetooth mode
    • WiFi
    • 16 GB memory
    Dual XBUS for Datrend test automation
    Power Supply
    • External AC adapter
    • Internal rechargeable Li-Ion batteries (for 10 hrs of simulation with full charge)
    Dimensions
    98mm x 275mm x 97mm (3.87" x 10.82" x 3.80")
    BP Unit (incl. A1 Base)
    Weight
    1080g (2.38lb) BP Unit (incl. A1 Base)
    200g (0.44lb) wireless tablet interface
    Environment
    15°C to 40°C, 10% to 90% RH, Indoor Use Only,
    Category II
  • vPad-BP Standard Accessories

    7500-146NIBP Hose Tubing Adapter Kit
    7500-505vPad-A1 Accessory Kit, NA
    7500-506vPad-A1 Accessory Kit, Europe
    7500-507vPad-A1 Accessory Kit, UK/Malaysia
    7500-508vPad-A1 Accessory Kit, AU/CN


    For more information, view the accessories list page