Specifications

vPad-BP™ Performance Specifications

Manufacturer Envelopes
  • GE Dinamap and Dinamap Pro
  • Critikon Dinamap Plus
  • Datascope Passport
  • Welch-Allyn Spot LXi and 52000
  • Fukuda Dynascope
  • Colin Press-Mate
  • …and more custom simulations
Pressure Units
mmHg, mbar, kPa, inH₂O, cmH₂O
BP Presets
Systolic/Diastolic (mmHg)
Adult Neonatal
255/195 150/100
200/150 120/80
150/100 100/65
120/80 80/50
100/65 60/30
80/50 35/15
60/30
BP Simulation:
Simulation type: oscillometric
Rate Range: 20 – 240 BPM
Accuracy:
  • ±0.25 BPM in sync mode
  • ±1 BPM otherwise
Amplitude:
  • 0 – 2 mL
  • 1.25 mmHg into 500ml cuff
Amplitude Range: 0 – 150%
Amplitude Accuracy: better than 0.5%
AutoSettings: unlimited, user definable
Envelope Shift:
± 50 mmHg max
Minimum Diastolic: 15 mmHg
Maximum Systolic: 275 mmHg
Manometer:
Pressure Range: 0.0 to 400.0 mmHg
Accuracy: ± 0.5 mmHg
Resolution: 0.1 mmHg
Regulated Pressure Source:
Pressure Range: 10.0 to 400.0 mmHg
Accuracy: ± 0.5 mmHg
Resolution: 0.1 mmHg
Leak Test:
Automatic/manual Inflation
Automatic Timer
Leak Test Time: 30 – 600 seconds (user defined)
Target Pressure: 20 to 400 mmHg
Range: 0 to 200 mmHg/min
User Definable Presets: 12
User Definable AutoSettings: unlimited
OverPressure Test:
Automatic/manual Inflation
Range: 20 – 400 mmHg
Release Time: 1-999 sec
User Definable Presets: 12
User Definable AutoSettings: unlimited
Standard Features / Accessories:
  • Autosequences
  • Unlimited User Defined Settings
  • vPad-A1 Power Base / Display
  • Universal Hose Adapter Kit
Communication / User Interface
via vPad-A1 Base Unit: Android 5″ tablet:

  • Touchscreen User Interface
  • Wired (USB) or Bluetooth mode
  • WiFi
  • 16 GB memory

Dual XBUS for Datrend test automation

Power Supply
  • External AC adapter
  • Internal rechargeable Li-Ion batteries (for 10 hrs of simulation with full charge)
Dimensions
98mm x 275mm x 97mm (3.87″ x 10.82″ x 3.80″)
BP Unit (incl. A1 Base)
Weight
1080g (2.38lb) BP Unit (incl. A1 Base)
200g (0.44lb) wireless tablet interface
Environment
15°C to 40°C, 10% to 90% RH, Indoor Use Only,
Category II