Specifications

AMPS-1 – Performance Specifications

ECG General:
Full 12-Lead ECG with 9 independent outputs for each signal lead referenced to RL.
Output Impedances: 500, 1000, 1500, & 2000 ohms to RL.
High Level Output: 0.5 V/mV of low level selection.
Amplitude Accuracy: ± 2% 2 Hz Square Wave (Lead II).
Normal Sinus Rhythm:
Rates: 30, 40, 60, 80, 100, 120, 140, 160, 180, 200, 220, 240, 260, 280, 300 BPM. Accuracy ± 1%.
Amplitudes (Lead II): 5mV, 4mV, 3mV, 2mV, 1mV, .5mV, .25mV, .1mV.
ST Segments: 16 total – 8 elevated & 8 depressed.
ST Segment Levels (Lead II): -0.8 mV to +0.8 mV in 0.1 mV steps on Lead II.
Axis Deviation: Normal (intermediate), horizontal, and vertical. Modifies baseline ECG during arrhythmias.
Neonatal Mode: ECG R wave width is reduced to 40ms.
ECG Performance Testing:
Square Wave: 2 Hz
Square Wave: 0.125 Hz
Pulse: 4.0 secs
Sine Waves: 0.05, 0.5, 1, 10, 25, 30, 40, 50, 60, and 100 Hz.
Triangle Wave: 2 Hz
R Wave Detector Test: 60 BPM haver-triangle wave with selectable ampl. and width.
Width: 8.0 ms to 200 ms (12 selections)
Amplitude (Lead II and V Leads): 5mV to 0.5mV.
Pacemaker:
Asynchronous
Demand with frequent sinus beat
Demand with occasional sinus beat
A-V sequential
Non-capture non-function
Rhythms: Async 75 BPM, Demand 1, Demand 2, AV Seq,
Non capture, Non function. – Pulse: -700 mV to +700 mV.
Accuracy: 10%
Width: 0.1, 0.2, 0.5, 1.0, 2.0 ms. Accuracy is 5%.
Pulse Polarity: Positive or negative.
Synchronization:
Accepts input during defib training.
RS-232 Interface: RS-232 interface to PC.
Defibrillator Training:
With two emergency scenarios and a cardioversion procedure, AMPS-1 can be used for basic defibrillator training.
Cardiac Output:
Built in Cardiac Output feature, activation optional
4 adjustable injectate temperature selections that are factory set and user adjustable
Baseline of 36, 37 and 38 degrees °C.
Selections for 2 and 20 degrees °C.
Cardiac Output Selections:
Faulty Injectate Curve
Left to Right Shunt Curve
C.O. of 3, 4, 4.5, 5, 5.5, 6, 6.5, 7 l/min
Cal Pulse: 1°C for 1 second
Cal Pulse: 1°C for 4 seconds
CC .561 for 2 degrees injectate
CC .608 for 20 degrees injectate
Temperature:
2 Temperature Channels
Electronically Switched Temperature of 35, 37, 38, 40, 42°C.
Probe Compatibility: 400 or 700 series YSI
Respiration:
Baseline Impedance: 500, 1000, 1500, 2000 ohms, LEADS I, II, III
Impedance Variations: 5, 4, 3, 2, 1, 0.5, 0.2, 0.1, 0.05, 0 Ω
Rates: 15 to 120 and 0 rpm for APNEA
Apnea Selections: 12, 22, 32 seconds, and continuous
Respiratory Effort (Inspiration/Expiration Ratio): 5/1, 4/1, 3/1 (normal), 2/1, 1/1.
Optional Fetal / Maternal / IUP Simulations:
Fetal heart rates: 60, 90, 120, 140, 150, 210, & 240 BPM
Trend Uniform, Early and Late deceleration
Maternal heart rate fixed at 80 BPM
Waveform: 12 lead ECG with complete p-qrs-t complex
Dynamic intrauterine pressure (iup) waveform: positive bell
shaped pressure curve
Peak pressure: 90 mmhg
Contraction duration: 90 sec
Pressure transducer sensitivity: 5 or 40 m v/v/mmhg
Input/output impedance: 300 ohms
Optional Mechanical Fetal Heart
Arrhythmia Selections:
Premature Beats Premature Atrial Contraction (PAC)
Nodal Premature Nodal Contraction (PNC)
Premature Ventricular Contraction (PVC)1
Left Ventricular Focus
PVC1 Early, Left Ventrical (LV) Focus
PVC1 R-on-T, Left Ventrical (LV) Focus
PVC2 Right Ventricular Focus
PVC2 Early, RV Focus
PVC2 R-on-T RV Focus
Multifocal PVCs
AED test waveforms:
  • • Atrial Fibrillation, Course Atrial fibrillation, Fine Ventricular Fibrillation, Course Ventricular Fibrillation
  • • Fine Asystole, Flatline Asystole, Random Baseline >0.1 mV
  • • Supraventricular Tachycardia Ventricular, Tachycardia @140,160 & 190BPM
  • • Torsades de Pointes @ 200 BPM NSR @ 60 BPM
Conduction Defects:
First Degree Heart Block Second Degree Heart Block
Third Degree Heart Block Right Bundle Branch Block
Left Bundle Branch Block Supraventricular Beats
Atrial Fibrillation (Coarse) Atrial Fibrillation (Fine)
Atrial Flutter Sinus Arrhythmia
Missed Beat (1 time event) Paroxysmal Atrial Tachycardia
Nodal Supraventricular Tachycardia
Ventricular Rhythm PVCs 6/Minute
PVCs 12/Minute PVCs 24/Minute
Frequent Multifocal Asystole
Pair PVCs (1 time event) Run 5 PVCs (1 time event)
Run 11 PVCs (1 time event) Ventricular Tachycardia
Ventricular Fibrillation (Coarse) Ventricular Fibrillation (Fine)
Bigeminy Trigeminy
4 Blood Pressure Channels:
Electrically Isolated Channels
Dynamic BP waveforms are synchronized with normal sinus rhythm rates and track arrhythmia selections.
Respiration artifact can be selected on blood pressure channels
Transducer Sensitivity: 5 or 40 μV/V/mmHg
Calibrated Rate: 80 BPM normal sinus rhythm
Static Levels BP1/2/3/4: -10, -5, 0, 20, 40, 50, 60, 80, 100, 120, 150, 160, 180, 200, 240, 320, 400 mmHg
Automatic Swan-Ganz (every 15 seconds)
Manual Swan-Ganz, changes each time Enter is selected

 

All specifications subject to change without notice.