Oxitest Plus7

oxitest_plus_7

Oxitest Plus7

SpO2 Simulator – with Nellcor® probe

  • High accuracy – Over 100 specific R-Curves
  • Easily test SPO2 meters with exclusive built-in “finger” sensor
  • Multiple pre-set tests
  • Add Sensitest for analysis of up to 5 unique probes

  • Overview
  • Features
  • Specifications
  • Accessories
  • Exclusive Nellcor-Puritan Bennett sensor port
    The versatile, handheld Oxitest Plus7 provides the most advanced technology available in the market today for verifying the proper operation and performance of pulse oximeter monitors and modules throughout the clinical environment.

    The Oxitest Plus 7 has an exclusive Nellcor-Puritan Bennett sensor port, which was developed in cooperation with Nellcor®. This Industry-exclusive feature allows correct selection of the instrument's simulation output to the specific R-Cal value of the attached Nellcor® probe.

    Compatibility beyond compare
    To further enhance your pulse oximeter testing requirements, Oxitest Plus 7 is compatible with Datrend's ES601 Plus and other automated test systems such as the Fluke® Biomedical medTester 5000C.

    Designed for compatibility with a wide variety of pulse oximeters, the Oxitest Plus 7 is the efficient, cost effective solution to your pulse oximeter testing requirements.
    • Easy adjustment of five preset oxygen saturation levels, six preset heart rates and four preset amplitudes.
    • Signal Artifact simulations
    • Preset patient parameters
    • Alarm Testing
    • Report printing
    • 2x20 character LCD display gives you all the information you need
    • Identifies the oximeter make and model
    • Status indicator of the test in progress with seven status messages
    • Rate
    • Saturation
    • Pulse Amplitude
  • Oxitest Plus7 - performance Specifications

    Saturation (SpO2):
    Range of adjustment is according to pulse oximeter manufacturer and/or model.
    Masimo, Mindray, Nellcor, and Nellcor OEM’s (e.g., Critikon; Protocol; Philips; etc.):
    Variable from 35% to 100% in steps of 1%
    SpO2 presets at 35, 70, 80, 90, 93, and 97%
    Selected EnviteC, Hewlett-Packard, Nihon Kohden, and Philips models:
    Variable from 55% to 100% in steps of 1%
    SpO2 presets at 55, 70, 80, 90, 93, and 97%
    All other manufacturers and models:
    Variable from 70% to 100% in steps of 1%
    SpO2 presets at 70, 80, 90, 93, and 97%
    SpO2 Accuracy:
    All supported oximeters except Datex and Invivo oximeters:
    55 to 100% SpO2:±1% at specified SpO2 presets
    35% SpO2:±2%
    Nellcor accuracy specified with either DS100A sensor, or with SpO2 Calibration selected via Nellcor Port
    Datex and Invivo oximeters:
    ±1% at 97, 93% SpO2
    ±2% at 90, 80, 70% SpO2
    Heart Rate:
    Variable from 20 to 250 BPM in 1 BPM steps
    Presets: 30, 60, 90, 120, 180, 240 BPM, Accuracy: ± 1 BPM
    Pulse Amplitude:
    Variable from zero (no blood flow) to 100% (normal adult pulse) in 1% steps.
    Presets: 100%, 30%, 10%, and 5%, Accuracy: ± 1%
    Signal Artifact:
    Four preset simulations:
    Movement
    Tapping (Spike artifact)
    Shivering (Tremor artifact)
    Shake Table (2.5Hz Sinewave)
    Auto Presets:Nine preset patient simulations. Default Auto Presets:
    Normal AdultTachycardiaLow Perfusion
    HypoxiaBardycardiaNo perfusion
    Movement ArtifactNeonatalTremor (Shivering Artifact)
    Auto Preset parameters may be user-programmed via the RS-232 port.
    Sensitest – Performance Specifications
    Fault Conditions Displayed:
    OPEN RED LED, INFRARED LED or PHOTODIODE LED or
    PHOTODIODE SHORT to ground or shield LED SHORT to
    PHOTODIODENellcor and Ohmeda connectors Standard.
    Options: Up to 3 sensor connectors can be added to the
    enclosure (total 5) from:
    Datex, Hewlett Packard, Novamatrix, Spacelabs, Invivo,
    Datascope, Criticare, Critikon, Sensormetics, PhysioControl,
    Simed and Nihon Koden.
    Alarm Tests:
    Five automated test sequences for determining oximeter alarm response time to Low SpO2 ; Low Heart Rate; High Heart Rate; Low Perfusion; and Motion Artifact conditions.
    Also determines time required for oximeter alarm to self-clear after test conditions are returned to normal.
    Alarm Test parameters may be user-programmed via the RS-232 port.
    Memory:
    Auto Preset and Alarm Test results saved in battery-backed, non-volatile Memory, along with up to 20 PASS/FAIL records from manually-conducted tests. Test results in Memory may be printed as a Test Report via optional external Printer, or may be downloaded via RS-232 port to personal computer or automated tester.
    User Interface:
    Display: LCD - 20 character x 2 line
    Keypad - 15 keys: PULSE OX, SPO (%), HEART RATE (preset), HR▲, HR▼, PULSE AMP (preset), PUL▲, PUL▼, PASS, FAIL, AUTO (preset), ALARM TEST, START/STOP (alarm test), ERASE (memory), PRINT
    Nellcor Port:
    Allows temporary connection of Nellcor pulse oximeter sensors for selection of sensor-specific SpO calibration curve. Intended for high-accuracy testing of all Nellcor disposable and re-usable sensors.
    Serial/Printer:
    Mechanical: 5 Pin MiniDIN / Electrical:
    Bi-directional RS-232.
    9600 baud, 8-N-1
    Power Supply:
    Battery: 7.2V NiCad rechargeable. Capacity: 1.4 Ah,
    Battery life: 40 hours continuous
    Standard Accessories:
    Power Adapter (Region-specific)
    Dimensions:
    5.12 in x 7.75 in x 1.5 in (13 cm x 19.68 cm x 3.8 cm)
    Weight:
    1.93 lbs (875g)
    Environment:
    15°C to 40°C, 10% to 90% RH, Indoor Use Only, Category II
     
    All specifications subject to change without notice.
  • Oxitest Accessories

    3140-060Cable, Molded MD5M TO RJ12, 6', Black (ES601 to Oxitest)
    3140-399Adaptor, USB To Serial DB9M
    6930-014Extended Calibration Certificate for Phase 3, DT-359
    7006-022Soft Carry Case For Oxitest (Black)
    7006-050Accessory Bag, 8.5" x 6" x 2.75" (Large)
    7006-053Soft Case, Small Multi-Product Bifold, 8" x 6" x 5-1/2"
    7100-248Cable Assy, OXT + & +7 RS232, 9 Pin (Oxitest to PC via RS-232)
    7500-099Kit, Upgrade to Oxitest V2.4
    7500-100SPS, 6W WPI, 100-240VAC, 12VDC/500mA, (Oxitest)
    7500-440Kit, Serial Adapter, Oxitest To PC Via USB (Oxitest to PC via USB)
    7500-449Oxitest Serial Printer Cable Kit for Epson LQ 300/310 Printer
    7500-4904-Blade Set For AC Input For 6W/15W WPI SPS (Oxitest) Contains NA, EU, UK & AU/CN

    For more information, view the accessories list page

Oxitest Plus7
Product Sheet
Oxitest Plus7
Support
Find Out More
Have A Dealer Contact Me
*
*
*
*