Specifications

vPad-353™ – Performance Specifications

Display:
10.1″ colour LCD (1280 x 800)
User Controls:
Capacitive touchscreen
Wired Connectivity:
Micro USB 2.0 Type B
(shareable with base unit)
XBUS Port (RJ11-6)
USB 2.0 Type A (host) – optional
Wireless Connectivity:
802.11 b/g/n
Bluetooth 2.1+ EDR
Modes of Operation:
Manual (standard)
Automatic (accessory option)
Power:
120 Volt power outlet: 90 to 132 V ac rms
47 to 63 Hz, 20 A max.
230 Volt power outlet: 180 to 264 V ac rms
47 to 63 Hz, 16 A max.
Weight:
5.5 lb (2.5 kg)
Size:
Base unit: 11.8 x 8.9 x 3.3 in.
(30 x 23 x 8.4 cm)
Tablet PC (detachable): 10.5 x 6.7 x 0.5 in.
(27 x 17 x 1.3 cm)
Test Standard Selections:
IEC 62353 AS 3551
ANSI/AAMI
ES1IEC60601-1
Voltage:
Mains voltage:
   Range:    90 to 264 V rms
   Accuracy:    ±(2 % of reading + 0.2 V)
External (point-to-point) voltage:
   Range:    0 to 300 V rms
   Accuracy:    ±(1 % FS + 0.2 V)
External (point-to-point) micropotential:
   Ranges:    0 to 199.9 mV rms200 to 1,999 mV rms2000 to 19,999 mV rms
   Accuracy:    ±(1 % of reading + 1 mV)
Protective Earth Resistance:
Method: Four-terminal, fully isolated
Test Current: 1A pulsed, 0.2A rms
Range: 0.000 to 2.000 Ω
Accuracy: ±(1 % of reading + 0.02 Ω)
Insulation Resistance:
Ranges: 0.5 to 5 MΩ5 to 50 MΩ50 to 999.9 MΩ
Range selection: Automatic
Accuracy:
   5 MΩ range    ±(1 % of reading + 0.1 MΩ)
   50 MΩ range    ±(2 % of reading + 0.2 MΩ)
   50 to 100 MΩ    ±(5 % of reading + 0.2 MΩ)
Test Voltage:
   Selections: 500 V or 250 V
   Accuracy: ±5% for 0 to 1 mA load
   Maximum Load Capacitance: 1 uF
Equipment Current:
Ranges 0 to 1.999 A ac rms2.00 to 19.99 A ac rms
Accuracy: ±(2 % reading + 0.2A)
Duty cycle 0 A to 10 A, continuous
10 A to 15 A, 7 min. on/3 min. off
15A to 20 A, 5 min. on/5 min. off
Equipment Power:
Range: 0 to 2400 watts
Accuracy: ±(5 % reading + 5W)
Equipment and Patient Leakage Tests:
Measurement: RMS
Method: Direct method
Patient Load: Per IEC 62353
Functional Group Selections: AP1; AP1 and AP2;
AP3 and AP4;
AP1 to AP3;
AP1 to AP4;
AP5 to AP10
AP1 to AP10
Chassis and Lead Leakage Tests:
Measurement: AC+DC (True-rms)
AC only
DC only
Patient Load Selection: AAMI ES1-1993
IEC 60601
Leakage Current Measurement:
Crest factor: <=3
Ranges: 0.0 to 199.9 μA
200 to 1,999 μA
2000 to 19,999 μA
Accuracy:
DC to 1 kHz ±(1 % of reading + 1μA)
1 to 100 kHz ±(2 % of reading + 1μA)
100 kHz to 1 MHz ±(5 % of reading + 1μA)
Isolation test voltage: 100 % ±5 % of AC supply
Calibration Test Points:
Resistance: 1 ± 0.01 Ω
Current: 100 ± 1 μA
ECG Performance Waveforms:
Output:
Amplitude 1 mV QRS into Lead II
Impedance 500 ohms
Accuracy:
Frequency ±1 %
Amplitude ±2%
Waveforms:
ECG Complex 30, 60, 120, 180, 240 and 300 BPM
Square wave 0.125 Hz, 2 Hz and 1 kHz
Pulse wave 63 msec, 30PPM and 60PPM
Triangle wave 2 Hz
Sine wave 0.5, 10, 40, 50, 60 and 100 Hz
CMRR tests SQR 2Hz & 1KHz, PUL 4s, SIN 0.5, 50 & 60Hz
Arryhthmias VFIB, AFIB, SVT, VTACH, PVC, and ASYS

 