AMPS-1

amps1

AMPS-1

Advanced Modular Patient Simulator

  • Full 12-lead ECG simulation with multiple outputs
  • Wide selection of arrhythmias
  • Temperature and respiration simulations
  • Optional Mechanical Fetal Heart and Cardiac Output Module
  • Portable, battery-operated

  • Overview
  • Features
  • Specifications
  • Accessories
  • Don’t let the size fool you 
    This compact Simulator packs a lot of punch in a small package. At only 7” X 4” X 1.25 inches, AMPS-1 fits comfortably in your hands and all leads and cables are conveniently fed from the top and bottom ends. Operate from the single 9 volt alkaline battery or with the optional battery eliminator.
     
    The unique modular design of AMPS-1 allows you to get just what you need as you need it. Upgrading and adding modules is easily accomplished by you at your facility. No need to send it to a repair facility just to add features.
  • The standard AMPS-1 is packed with...
    • 12 lead ECG simulation with 9 independent outputs for each signal lead
    • 16 total ST Segments: 8 elevated and 8 depressed
    • Axis Deviation: Normal (intermediate), horizontal, and vertical. (Modifies baseline ECG during arrhythmias)
    • Neonatal Mode: ECG R wave width is reduced to 40ms
    • ECG Performance Testing
    • 52 Arrhythmia selections
    • Temperature and Respiration Simulations
    • Pacer simulations
    • Defibrillator training
    • Remote Control via RS-232

    As your workload and budget dictate, easily add...
    • 2 or 4 Electrically Isolated BP Channels, including Swan-Ganz simulation
    • Cardiac output
    • Mechanical Fetal Heart
    • Fetal, Maternal and IUP simulations
  • AMPS-1 – Performance Specifications

    ECG General:
    Full 12-Lead ECG with 9 independent outputs for each signal lead referenced to RL.
    Output Impedances:500, 1000, 1500, & 2000 ohms to RL.
    High Level Output:0.5 V/mV of low level selection.
    Amplitude Accuracy:± 2% 2 Hz Square Wave (Lead II).
    Normal Sinus Rhythm:
    Rates:30, 40, 60, 80, 100, 120, 140, 160, 180, 200, 220, 240, 260, 280, 300 BPM. Accuracy ± 1%.
    Amplitudes (Lead II):5mV, 4mV, 3mV, 2mV, 1mV, .5mV, .25mV, .1mV.
    ST Segments:16 total – 8 elevated & 8 depressed.
    ST Segment Levels (Lead II):-0.8 mV to +0.8 mV in 0.1 mV steps on Lead II.
    Axis Deviation:Normal (intermediate), horizontal, and vertical. Modifies baseline ECG during arrhythmias.
    Neonatal Mode:ECG R wave width is reduced to 40ms.
    ECG Performance Testing:
    Square Wave:2 Hz
    Square Wave:0.125 Hz
    Pulse:4.0 secs
    Sine Waves:0.05, 0.5, 1, 10, 25, 30, 40, 50, 60, and 100 Hz.
    Triangle Wave:2 Hz
    R Wave Detector Test:60 BPM haver-triangle wave with selectable ampl. and width.
    Width:8.0 ms to 200 ms (12 selections)
    Amplitude (Lead II and V Leads):5mV to 0.5mV.
    Pacemaker:
    Asynchronous
    Demand with frequent sinus beat
    Demand with occasional sinus beat
    A-V sequential
    Non-capture non-function
    Rhythms:Async 75 BPM, Demand 1, Demand 2, AV Seq,
    Non capture, Non function. - Pulse:-700 mV to +700 mV.
    Accuracy:10%
    Width:0.1, 0.2, 0.5, 1.0, 2.0 ms. Accuracy is 5%.
    Pulse Polarity:Positive or negative.
    Synchronization:
    Accepts input during defib training.
    RS-232 Interface: RS-232 interface to PC.
    Defibrillator Training:
    With two emergency scenarios and a cardioversion procedure, AMPS-1 can be used for basic defibrillator training.
    Cardiac Output:
    Built in Cardiac Output feature, activation optional
    4 adjustable injectate temperature selections that are factory set and user adjustable
    Baseline of 36, 37 and 38 degrees °C.
    Selections for 2 and 20 degrees °C.
    Cardiac Output Selections:
    Faulty Injectate Curve
    Left to Right Shunt Curve
    C.O. of 3, 4, 4.5, 5, 5.5, 6, 6.5, 7 l/min
    Cal Pulse: 1°C for 1 second
    Cal Pulse: 1°C for 4 seconds
    CC .561 for 2 degrees injectate
    CC .608 for 20 degrees injectate
    Temperature:
    2 Temperature Channels
    Electronically Switched Temperature of 35, 37, 38, 40, 42°C.
    Probe Compatibility: 400 or 700 series YSI
    Respiration:
    Baseline Impedance:500, 1000, 1500, 2000 ohms, LEADS I, II, III
    Impedance Variations:5, 4, 3, 2, 1, 0.5, 0.2, 0.1, 0.05, 0 Ω
    Rates:15 to 120 and 0 rpm for APNEA
    Apnea Selections:12, 22, 32 seconds, and continuous
    Respiratory Effort (Inspiration/Expiration Ratio):5/1, 4/1, 3/1 (normal), 2/1, 1/1.
    Optional Fetal / Maternal / IUP Simulations:
    Fetal heart rates:60, 90, 120, 140, 150, 210, & 240 BPM
    Trend Uniform, Early and Late deceleration
    Maternal heart rate fixed at 80 BPM
    Waveform:12 lead ECG with complete p-qrs-t complex
    Dynamic intrauterine pressure (iup) waveform: positive bell
    shaped pressure curve
    Peak pressure:90 mmhg
    Contraction duration:90 sec
    Pressure transducer sensitivity:5 or 40 m v/v/mmhg
    Input/output impedance:300 ohms
    Optional Mechanical Fetal Heart
    Arrhythmia Selections:
    Premature Beats Premature Atrial Contraction (PAC)
    Nodal Premature Nodal Contraction (PNC)
    Premature Ventricular Contraction (PVC)1
    Left Ventricular Focus
    PVC1 Early, Left Ventrical (LV) Focus
    PVC1 R-on-T, Left Ventrical (LV) Focus
    PVC2 Right Ventricular Focus
    PVC2 Early, RV Focus
    PVC2 R-on-T RV Focus
    Multifocal PVCs
    AED test waveforms:
    • • Atrial Fibrillation, Course Atrial fibrillation, Fine Ventricular Fibrillation, Course Ventricular Fibrillation
    • • Fine Asystole, Flatline Asystole, Random Baseline >0.1 mV
    • • Supraventricular Tachycardia Ventricular, Tachycardia @140,160 & 190BPM
    • • Torsades de Pointes @ 200 BPM NSR @ 60 BPM
    Conduction Defects:
    First Degree Heart BlockSecond Degree Heart Block
    Third Degree Heart BlockRight Bundle Branch Block
    Left Bundle Branch BlockSupraventricular Beats
    Atrial Fibrillation (Coarse)Atrial Fibrillation (Fine)
    Atrial FlutterSinus Arrhythmia
    Missed Beat (1 time event)Paroxysmal Atrial Tachycardia
    NodalSupraventricular Tachycardia
    Ventricular RhythmPVCs 6/Minute
    PVCs 12/MinutePVCs 24/Minute
    Frequent MultifocalAsystole
    Pair PVCs (1 time event)Run 5 PVCs (1 time event)
    Run 11 PVCs (1 time event)Ventricular Tachycardia
    Ventricular Fibrillation (Coarse)Ventricular Fibrillation (Fine)
    BigeminyTrigeminy
    4 Blood Pressure Channels:
    Electrically Isolated Channels
    Dynamic BP waveforms are synchronized with normal sinus rhythm rates and track arrhythmia selections.
    Respiration artifact can be selected on blood pressure channels
    Transducer Sensitivity:5 or 40 μV/V/mmHg
    Calibrated Rate:80 BPM normal sinus rhythm
    Static Levels BP1/2/3/4:-10, -5, 0, 20, 40, 50, 60, 80, 100, 120, 150, 160, 180, 200, 240, 320, 400 mmHg
    Automatic Swan-Ganz (every 15 seconds)
    Manual Swan-Ganz, changes each time Enter is selected
     
    All specifications subject to change without notice.
  • AMPS-1 Accessories

    3000-445LPS, 120VAC / 9VDC,0.5A, EIAJ-3, C+ GND, NA for AMPS-1/MPS-2/FMS-3
    3000-446LPS, 220VAC / 9VDC,0.5A, EIAJ-3, C+ GND,EU for AMPS-1/MPS-2/FMS-3
    3000-447LPS, 240VAC / 9VDC,0.6A, EIAJ-3, C+ GND, U.K. for AMPS-1/MPS-2/FMS-3
    3140-429Cable, RJ12(6P6C) - RJ12), Tester Interface, 7' (ES601 to AMPS-1)
    6930-011Extended Calibration Certificate for AMPS-1, DT-357
    6960-100AMPS-1, 2 Electrically Isolated BP Channels (AMPS-1 software option)
    6960-101AMPS-1, 4 Electrically Isolated BP Channels (AMPS-1 software option)
    6960-102AMPS-1, Cardiac Output (AMPS-1 software option) requires 6960-100 or 6960-101
    6960-103AMPS-1, Fetal, Maternal & IUP (AMPS-1 software option) requires 6960-100 or 6960-101
    7006-053Soft Case, Small Multi-Product Bifold, 8" x 6" x 5-1/2"
    7200-098Cable Assy, Unterminated IBP Cable
    7200-239 Cable Assy, Colin A308 6Pin IBP Cable
    7200-454YSI 400 Temp Cable (6 Pin Mini-Din - 1/4" Mono Phono Plug)
    7200-455YSI 700 Temp Cable (6 Pin Mini-Din - 1/4" Stereo Phono Plug)
    7200-463 Invasive BP Cable, Datascope, 6 Socket
    7200-464 Invasive BP Cable, Datex-Ohmeda, 10 Socket
    7200-466 Invasive BP Cable, GE/Marquette, 11 Pin
    7200-468 Invasive BP Cable, HP/Invivo/MDE, 6 Pin
    7200-469 Invasive BP Cable, HP/Philips, 12 Pin
    7200-470 Invasive BP Cable, Marquette, 7 Pin
    7200-471 Invasive BP Cable, Nihon Khoden, 5 Pin
    7200-477 Invasive BP Cable, Fukuda Denshi, 12 Pin
    7200-479 Invasive BP Cable, Criticare Compass
    7200-523IBP Adapter Cable, Siemens 10 Pin - 7 Pin
    7200-524 IBP Cable Siemens 10 Pin
    7200-530Temperature Cable, Philips / HP Simulator Cable (AMPS-1/MPS-2)
    7500-435Kit, Serial Adapter, ES601 -> P3/Solo, AMPS-1 -> PC (RJ12 -> DB9F) (ES601 to Solo/Phase3)
    8000-454Cardiac Output Adapter requires 6960-102
    8000-455Mechanical Fetal Heart Adapter (LPS Required)

    For more information, view the accessories list page

AMPS-1
Product Sheet